top of page

Specifications
Model Name: FXRD-1717NA / FXRD-1717NBFXRD-1717NAW / FXRD-1717NBW
Technology: a-Si TFT
Scintillator: Csl / Gadox
Pixel Pitch: 140 ㎛
Spatial Resolution: 3.5 lp/㎜
Pixel Matrix: 3072 x 3072 pixels
Image Size: 17" x 17" (43.0 ㎝ × 43.0 ㎝)
Grayscale: 16 bits
Image Acquisition Time: Wired: 1.5 s / Wireless: 3 s
Recommended Cycle Time: 15 s
Data Interface: Gigabit Ethernet / IEEE 802.11n / ac
X-ray Generator Interface: DR trigger Mode / AED Mode
Dimensions: 46.0 ㎝ × 46.0 ㎝ × 1.5 ㎝
Weight: Wired : 4.2 kg / Wireless: 4.5 kg (including battery)
Battery: Lithium Ion 3400mAh x 2 / 6 h (capturing)
Dust and Water Resistant: IP53
X-ray Voltage Range: 40 - 150 kVp
Operating Environment: 10℃ to 35℃ / 30 % to 85 % RH (non-condensing)
Power Consumption: DC 24 V, 1.0 A